Comparison of pulmonary vein isolation between two commercially available cryoballoon systems

Abstract

Background

Pulmonary vein isolation (PVI) using cryoballoon (CB) ablation has comparable efficacy and safety to the gold standard of radiofrequency ablation in the treatment of symptomatic atrial fibrillation (AF). Initial randomized control trials were performed using Arctic Front Advance Pro™ (AFr) (Medtronic, Dublin, Ireland) CB system. Novel CB systems have recently become available, including the POLARx™ (Px) (Boston Scientific, Marlborough, Massachusetts, USA) system. We aimed to compare PVI using the Px and the AFr CB systems in our patient population in terms of efficacy, safety and procedure characteristics in a routine clinical setting.

Methods

We performed a retrospective analysis of our internal AF ablation registry, containing 452 consecutive patients (pts) that underwent first procedure cryo-PVI for symptomatic AF. Primary endpoints were AF recurrence after 3 and 12 months, complication rate, procedure duration, fluoroscopy time and fluoroscopy dose. Secondary endpoints were minimal freeze temperature, time to isolation (TTI) and temperature at TTI for each of the pulmonary veins as well as minimal esophageal temperature during the procedure.

Results

The primary efficacy endpoints of AF recurrence after 3 and 12 months were similar between the AFr and the Px systems (25.5% vs 21.3%, p = 0.416 and 22.2% vs 20.6%, p = 0.794, respectively). Complication rates were similar (3.9% vs 6.8%, p = 0.18) between groups and consisted mostly of mild vascular complications. The AFr group showed a significantly shorter procedural duration (68 (55–77) vs 73 (60–80) min, p = 0.002), and lower fluoroscopy dose compared to the Px system. Fluoroscopy times remained similar, however. Minimal freeze temperatures and temperatures at time of isolation were significantly lower in the Px group. However, the time to isolation and minimal procedural esophageal temperature were similar in both groups.

Conclusion

PVI using the AFr and the Px systems showed comparable safety and efficacy. Procedural times were longer for the Px system. The Px system showed lower freeze temperature measurements but seemed to have a comparable biological effect.

Background

Initial randomized controlled trials (RCTs) showed comparable efficacy and safety and significantly shorter procedural times for pulmonary vein isolation (PVI) using cryoballoon ablation (CBA) compared to the gold standard of radiofrequency ablation (RFA) in the treatment of symptomatic atrial fibrillation (AF) [1,2,3].

A recent metanalysis based on 14 randomized controlled trials (RCTs) and 34 observational studies showed a reduced incidence of AF recurrence for CBA compared with RFA, a difference that was not dependent on study design [4]. Although complication rates were higher in the CBA group, the difference was driven by (mostly transient) phrenic nerve palsy [4]. The rates of pericardial effusion, cardiac tamponade and vascular complications were lower in the CBA group [4]. In conclusion, CBA presents a favorable method for first procedure pulmonary vein isolation, with possible improvements to be made in prevention of phrenic nerve palsy.

Initial RCTs were performed using Arctic Front Advance ™ (Medtronic, Dublin, Ireland) CB system [1]. Meanwhile, the fourth-generation Arctic Front Advance Pro™ (AFr) system is being routinely used, allowing for better real-time capture of PV potentials due to a shorter tip size [5].

Novel CB systems have recently become available, including the POLARx™ (Px) (Boston Scientific, Marlborough, Massachusetts, USA) system [6]. Several observational studies showed comparable efficacy and safety between the AFr and the Px systems [7,8,9,10,11]. Randomized studies are underway [12].

We aimed to compare PVI using the Px and the AFr CB systems in our patient population in terms of efficacy, safety and procedure characteristics in a routine clinical setting. The rationale is providing data specific for our population and setting, as well as data on rare complications like phrenic nerve palsy and cardiac tamponade which can be pooled by future research efforts.

Methods

We performed a retrospective analysis of our internal AF ablation registry, containing 452 consecutive patients that underwent first procedure cryo-PVI for symptomatic AF from October 2020 to February 2024. The new Px CB system was acquired by our center in October 2020. The AFr System has been in routine use for more than 5 years. The comparison of characteristics of the two systems can be found in Table 1. Esophageal temperature measurement and phrenic nerve stimulation for prevention of phrenic nerve palsy were performed in all patients.

Table 1 Balloon characteristics

Procedural sedation and analgesia were performed using a combination of propofol, fentanyl and midazolam. A steerable decapolar catheter was placed in the coronary sinus by femoral access and was used as a landmark to perform a single trans-septal puncture (TSP). Routine administration of unfractionated heparin was performed immediately after TSP with an initial bolus of 160 IU/kg and with a target activated clotting time of > 300 s.

During ablation of the right pulmonary veins, the decapolar catheter was positioned in the superior vena cava to perform phrenic nerve stimulation. Diaphragmatic action was monitored using modified ECG leads in both systems, and with an additional integrated diaphragm movement sensor using the Px system.

AF recurrence was defined as a documented episode of AF or atrial tachycardia (AT) > 30 s. Patients underwent follow-ups at 3 months and 12 months, either as part of a prescheduled visit or via telephone call to the primary care provider or to the patient. Recurrence during the first 3 months after ablation was excluded at 12 months of follow-up. Most of the patients received at least one Holter ECG over at least 24 h during each follow-up period or AF recurrence was determined using an implanted device interrogation. AF recurrence in the remaining patients was assessed based on symptoms and occasional ECGs.

Primary endpoints were AF recurrence after 3 and 12 months, complication rate, procedure duration, fluoroscopy time and fluoroscopy dose. Secondary endpoints were minimal freeze temperature, time to isolation (TTI) and temperature at TTI for each of the pulmonary veins as well as minimal esophageal temperature during the procedure.

Statistical analysis

Continuous variables are presented as mean ± standard deviation or median and interquartile range as appropriate and categorical variables are summarized as absolute and relative frequencies. The Kolmogorov–Smirnov test was used to test normality. Continuous variables were compared using the Student’s t-test or the Mann–Whitney U test. Pearson's chi‐square test was used for categorical variables. A two-sided p value of < 0.05 was considered statistically significant.

The data for the procedural times of the Px system were analyzed to evaluate the learning curve. The dataset was divided into two groups: early procedures and late procedures, with the split point set at the midpoint of the dataset. A t-test was performed to compare the mean procedural times between the early and late groups.

Analyses were performed using the SPSS software, version 26.0 (IBM Corporation, Armonk, NY, USA).

Results

Baseline characteristics

We included 452 consecutive patients who underwent CBA from October 2020 to February 2024. Of the initial population (n = 452, 61% male, mean age 65.0 ± 12.7 years, BMI 27.8 ± 4.5 kg/m2, LA 38.5 ± 5.9 mm), 51% had paroxysmal AF (Table 2). A total of 306 Patients received CBA with the AFr system versus 146 patients with the Px System. There were significantly more patients with paroxysmal AF in the AFr group (55.9% vs 42.1%, p = 0.006), and the mean LA diameter was slightly larger in the Px group (37.9 ± 5.7 vs 39.7 ± 6.3, p = 0.005) (Table 2). At discharge, 26% of patients were on class Ic or III antiarrhythmic drugs and 75% were on beta blockers.

Table 2 Baseline characteristics

Primary endpoints (Table 3)

Table 3 Primary endpoints

Acute PVI of all veins was achieved in all patients. The primary efficacy endpoints of AF recurrence after 3 and 12 months were similar between the AFr and the Px systems (25.5% vs 21.3%, p = 0.416 and 22.2% vs 20.6%, p = 0.794, respectively) (Fig. 1) (Table 3). Complication rates were similar (3.9% vs 6.8%, p = 0.18) between groups and consisted mostly of mild haematoma (0.9% vs 3.5%, p = 0.066) (Table 3). One patient in the Px group needed vascular surgery for arteriovenous fistula (0.7%). There was one case (0.7%) of suspected right coronary artery air embolism without sequelae in the Px group, as well as one case of stroke (0.3%) and one case of transient ischemic attack (0.3%), both occurring postprocedurally, in the AFr group. Phrenic nerve palsy occurred in 2.4% of patients in the AFr group compared to 2.1% in the Px group (p = 0.799).

Fig. 1
figure 1

Primary efficacy endpoints

The AFr group showed a significantly shorter procedural duration (68 (55–77) vs 73 (60–80) min, p = 0.002) and lower fluoroscopy dose (2209 (1298–4600) vs 2997 (1584–6938) mGycm2, p = 0.021) compared to the Px system. Fluoroscopy times remained similar (10.3 (8.1–14.2) vs 10.6 (8.1–14.7) min, p = 0.851), however (Table 3). The analysis of the learning curve for the Px system showed no significant difference in procedural duration between the early and the late procedures (p = 0.417) (Fig. 2).

Fig. 2
figure 2

Learning curve for the Px procedure

Secondary endpoints (Tables 4, 5)

Table 4 Secondary endpoints—freeze parameters
Table 5 Secondary endpoints—minimal esophageal temperature

Minimal freeze temperatures (− 52, − 48, − 54 and − 51 °C vs − 56.5, − 53, − 60, and − 57 °C for LSPV, LIPV, RSPV and RIPV, respectively, p < 0.001 for all PVs) and temperatures at time of isolation (− 40, − 37, − 39.5 and − 40 vs − 48.5, − 42.5, − 42.5 and − 43 °C for LSPV, LIPV, RSPV and RIPV, respectively, p < 0.001 for all PVs) were significantly lower in the Px group (Table 4). However, the time to isolation (49 (37–64), 39 (25–74), 40 (28–66) and 52.5 (36–90) vs 60 (38–74), 56.5 (35–87), 44.5 (25–79), 47.5 (31–100) s, p = 0.819, 0.218, 0.424, and 0.062, for LSPV, LIPV, RSPV and RIPV, respectively) and minimal procedural esophageal temperature (26.2 (17.0–29.8) vs 29.8 (14.6–33.7), p = 0.465) were similar in both groups.

Discussion

This study aimed to compare efficacy, safety and procedure characteristics of PVI using the Px and the AFr CB systems in our patient population. We found similar efficacy in terms of AF recurrence after 3 and 12 months. The complication rate was comparable between the groups. The numerical difference in complication rate was driven by higher rates of mild hematoma in the Px group, which could be explained by a slight difference in sheath size (15.9 F for Px vs 15 F for AFr).

Rates of phrenic nerve palsy were comparable between the two systems. We observed one case of RCA air embolism without sequelae using the Px system as well as one case of stroke (0.5%) and one case of transient ischemic attack (0.5%), both occurring postprocedurally, in the AFr group—the event rate is too low to reach statistical significance. Larger studies and pooled analyses might elucidate whether there is a difference in rare complications between the two technologies.

Our study is largely comparable to previously published data: past observational studies showed comparable efficacy between the AFr and the Px systems [7,8,9,10]. One small observational study showed a trend toward a higher rate of phrenic nerve palsy in the Px group (15% vs 7%, p = 0.05) [11], whereas other studies showed similar safety [7,8,9,10]. Larger studies or pooled data analysis will be necessary to determine whether there are differences in very rare complications like TIA, stroke, cardiac tamponade or atrioesophageal fistula.

The Px system showed lower minimal freeze temperatures as well as temperatures at TTI. However, the TTI and minimal esophageal temperature were similar between systems.

Our results are consistent with earlier research—in small observational studies, minimal freeze temperatures and temperatures at time to isolation (TTI) were significantly lower for the Px system, whereas time to isolation (TTI) was similar [7, 8, 11]. This points to a difference in temperature measurement due to differences in the design of the systems, as opposed to a stronger biological effect of the Px system.

The procedural times were longer in the Px group. This is consistent with earlier studies [7, 8] and might be explained by inexperience using the new system or by a difference in handling of the systems. The analysis of the learning curve for the Px system showed no significant difference in procedural duration between the early and the late procedures (Fig. 2). However, the study might be underpowered to find a difference in this highly variable endpoint.

Study limitations

This was a small observational retrospective single center study with associated limitations. Holter ECG or device interrogation data were available for 55.1% patients. In the remaining patients, recurrence was assessed solely on the basis of symptoms and occasional ECGs. This might have led to some asymptomatic AF recurrence events being missed, but should not alter the relative procedural success between both systems.

Conclusion

PVI using the AFr and the Px systems showed comparable safety and efficacy. Procedural times were longer for the Px system. The Px system showed lower freeze temperature measurements but seemed to have a comparable biological effect.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

References:

  1. Kuck KH, Brugada J, Fürnkranz A, Metzner A, Ouyang F, Chun KRJ, et al. Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl J Med. 2016;374(23):2235–45.

    Article  PubMed  Google Scholar 

  2. Luik A, Radzewitz A, Kieser M, Walter M, Bramlage P, Hörmann P, et al. Cryoballoon versus open irrigated radiofrequency ablation in patients with paroxysmal atrial fibrillation: the prospective, randomized, controlled, Noninferiority FreezeAF Study. Circulation. 2015;132(14):1311–9.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Andrade JG, Champagne J, Dubuc M, Deyell MW, Verma A, Macle L, et al. Cryoballoon or radiofrequency ablation for atrial fibrillation assessed by continuous monitoring: a randomized clinical trial. Circulation. 2019;140(22):1779–88.

    Article  PubMed  Google Scholar 

  4. Fortuni F, Casula M, Sanzo A, Angelini F, Cornara S, Somaschini A, et al. Meta-Analysis comparing cryoballoon versus radiofrequency as first ablation procedure for atrial fibrillation. Am J Cardiol. 2020;125(8):1170–9.

    Article  PubMed  Google Scholar 

  5. Moltrasio M, Sicuso R, Fassini GM, Riva SI, Tundo F, Dello Russo A, et al. Acute outcome after a single cryoballoon ablation: comparison between Arctic Front Advance and Arctic Front Advance PRO. Pacing Clin Electrophysiol. 2019;42(7):890–6.

    Article  PubMed  Google Scholar 

  6. Anic A, Lever N, Martin A, Breskovic T, Sulkin MS, Duffy E, et al. Acute safety, efficacy, and advantages of a novel cryoballoon ablation system for pulmonary vein isolation in patients with paroxysmal atrial fibrillation: initial clinical experience. EP Europace. 2021;23(8):1237–43.

    Article  Google Scholar 

  7. Yap S, Anic A, Breskovic T, Haas A, Bhagwandien RE, Jurisic Z, et al. Comparison of procedural efficacy and biophysical parameters between two competing cryoballoon technologies for pulmonary vein isolation: insights from an initial multicenter experience. J Cardiovasc Electrophysiol. 2021;32(3):580–7.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Knecht S, Sticherling C, Roten L, Badertscher P, Chollet L, Küffer T, et al. Technical and procedural comparison of two different cryoballoon ablation systems in patients with atrial fibrillation. J Interv Card Electrophysiol. 2021. https://doi.org/10.1007/s10840-021-01035-6.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Knecht S, Sticherling C, Roten L, Badertscher P, Krisai P, Chollet L, et al. Efficacy and safety of a novel cryoballoon ablation system: multicentre comparison of 1-year outcome. EP Europace. 2022;24(12):1926–32.

    Article  Google Scholar 

  10. Menger V, Frick M, Sharif-Yakan A, Emrani M, Zink MD, Napp A, et al. Procedural performance between two cryoballoon systems for ablation of atrial fibrillation depends on pulmonary vein anatomy. J Arrhythmia. 2023;39(3):341–51.

    Article  Google Scholar 

  11. Tanese N, Almorad A, Pannone L, Defaye P, Jacob S, Kilani MB, et al. Outcomes after cryoballoon ablation of paroxysmal atrial fibrillation with the PolarX or the Arctic Front Advance Pro: a prospective multicentre experience. EP Europace. 2023;25(3):873–9.

    Article  PubMed  Google Scholar 

  12. Maurhofer J, Kueffer T, Knecht S, Madaffari A, Badertscher P, Seiler J, et al. Comparison of the PolarX and the Arctic Front cryoballoon for pulmonary vein isolation in patients with symptomatic paroxysmal atrial fibrillation (COMPARE CRYO)—study protocol for a randomized controlled trial. Contemp Clin Trials. 2023;134:107341.

    Article  PubMed  Google Scholar 

Download references

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

J-H van den B, JW, CS, J-H S, SD and KF were involved in data collection. KF and DS performed data analysis and interpretation and wrote the manuscript. AS and JL contributed to data interpretation. All authors contributed to the writing of the final version of the manuscript, read and approved the final version.

Corresponding author

Correspondence to Karlo Filipovic.

Ethics declarations

Ethics approval and consent to participate

Ethical approval was obtained from the local ethical committee.

Consent for publication

Not applicable (none required from the local ethical committee).

Competing interests

D. Steven reports receiving grants from Biosense Webster and Abbott as well as lecture fees from Medtronic, Pfizer, Biosense and Abbott. J. Lüker reports having received lecture fees from Boston Scientific, Johnson&Johnson and Abbott. J. Wörmann reports having received lecture fees from Abbott and Boston Scientific and educational fees from Boston Scientific and Johnson&Johnson. Other authors report having no conflicts of interest.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Filipovic, K., Sultan, A., Lüker, J. et al. Comparison of pulmonary vein isolation between two commercially available cryoballoon systems. Int J Arrhythm 25, 17 (2024). https://doi.org/10.1186/s42444-024-00124-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s42444-024-00124-7